minolta spectrophotometer
CM-3600d / 3610d
Spectrophotometer (spectral type, side port / bottom port) . Horizontal Spectrophotometer CM-3600d and Vertical Spectrophotometer CM-3610d are equipped with Konica Minolta Innovative Optical System technology
Principal Uses
Reflected and transmitted color measurement , Research, development, and quality control in a wide range of industries. Universities, testing laboratories, and examination bodies.
Illuminating/viewing system | Reflectance: d/8 (diffuse illumination/8° viewing angle), capable of simultaneous SCI (specular component included)/SCE (specular component excluded) measurement. Conforms to CIE No. 15, ISO 7724/1, ASTM E-1164, DIN 5033 Teil 7, and JIS Z8722 Condition C standards. Transmittance: d/0 (diffuse illumination/0° viewing angle). Conforms to CIE No. 15, ASTM E-1164, DIN 5033 Teil 7 standards |
---|---|
Detector | Dual 40-element silicon photodiode arrays |
Spectral separation device | Diffraction grating |
Wavelength range | 360 to 740nm |
Wavelength pitch | 10nm |
Half bandwidth | Approx. 10nm |
Photometric range | 0 to 200%; Display resolution: 0.01% |
Sphere size | Φ152mm |
Light source | Four pulsed xenon lamps |
Measurement time | Approximately 1.5s (until data output) |
Minimum interval between measurements | Approx. 4 seconds when SCI/SCE measured |
Measurement/illumination areas | Reflectance: Changeable between LAV, MAV, and SAV LAV: Φ25.4mm/Φ30m MAV: Φ8mm/Φ11m SAV: Φ4mm/Φ7mm Transmittance: Approx. 17mm |
Repeatability | Spectral reflectance: Standard deviation within 0.1% Colorimetric values: Standard deviation within ΔE*ab 0.02 |
Inter instrument agreement | Mean ΔE*ab 0.15 (LAV/SCI) Average for 12 BCRA Series II color tiles compared to values measured with master body |
Temperature dependence | Spectral reflectance: Within ± 0.10%/°C Color difference: Within ΔE*ab 0.05/°C |
UV adjustment | Instantaneous numerical adjustment |
UV cut filters | 400nm cutoff and 420nm cutoff |
Transmittance chamber | Width: 133mm, Depth: Approx. 50mm; Measurement diameter: Approx. 17mm Transmission sample holder (optional accessory): Sample holder; for both plate-shaped and liquid samples (removable) |
Interface | RS-232C, D-sub 9-pin (female) terminal |
Power | AC 120V/230V 50/60Hz (using included AC adapter) |
Operating temperature/humidity range | 13 to 33°C, relative humidity 80% or less (at 33°C) with no condensation |
Storage temperature/humidity range | 0 to 40°C, relative humidity 80% or less (at 33°C) with no condensation |
Dimensions (W × H × D) | CM-3600d: 244 × 208 × 378mm CM-3610d: 300 × 597 × 315mm |
Weight | CM-3600d: 12kg CM-3610d: 16.5kg |
Standard accessories | White Calibration Plate, Target Mask (Φ4mm), Target Mask (Φ8mm), Target Mask (Φ25.4mm), Zero Calibration Box, AC Adapter, Dust Cover, Accessory Case, Unit Driver, RS-232C Cable (9-pin, 2m) |
Optional accessories | Color Data Software SpectraMagic™ NX, Transmittance Specimen Holder, Cell (2mm)/(10mm)/(20mm), Transmittance Zero Calibration Plate, RS-232C Cable (IBM PC/AT; 5m)/(IBM PS/2; 2m)/(IBM PS/2; 5m) |